Ensuring Accurate Blood Pressure Monitoring for Bariatric Patients

Principal Investigator: Margo A. Halm, RN, PhD, CCRN, APRN-BC
Co-Principal Investigator/Collaborator: Becky Braden
Organization: United Hospital

Abstract

Problem

Over 60% of the U.S. population is overweight or obese. As the bariatric population grows, nurses face unique challenges to ensure clinically effective care. Our Nursing Research Council (NRC) investigated a concern that BP readings that were relied on for treatment were frequently highly variable. Indeed, multiple cuff sizes and sites of BP assessment, as well as inconsistent practices between caregivers, were discovered.

Evidence

The synthesis of evidence revealed correct cuff size is paramount for accurate BP. If a too narrow or wide cuff is used (i.e., regular vs. thigh), BP will be over- or underestimated, respectively. More evidence supported using disposable cuffs to eliminate cross-contamination.

Strategy

The NRC designed an evidence-based practice (EBP) protocol to assess if measuring and selecting appropriate cuffs reduced BP variation in bariatric surgical patients.

Practice Change

Preoperative nurses measured the circumference of the patients’ upper arm and selected the appropriate disposable cuff to be used throughout the entire postoperative stay.

Evaluation

Statistical process control methods evaluated BP variability, and observation audits assessed staff compliance with the protocol.

Results

Control charts did not show less “out of control” BP readings, suggesting other factors were affecting BP (e.g., pain, compliance with cuff selection, or consistent caregiver use). Observation audits revealed over 85% of nurses followed the EBP protocol. Nurse and physician satisfaction showed that 60% were satisfied with the bariatric BP protocol however, only 46% believed it increased clinical reliability of BP readings.

Recommendations

While the EBP project did not reduce BP variation, staff learned how critical an appropriately-sized and consistently used cuff is for bariatric patients over the course of hospitalization. Consistent practice allows for trending of BP’s which may be more useful to guide treatment until specially designed conical-shaped BP cuffs for the bariatric population are tested and become readily available.

Bibliography

  • Base-Smith, V. (1996). Non–disposable sphygmomanometer cuffs harbor frequent bacterial colonization and significant contamination by organic and inorganic matter. J American Association of Nurse Anesthetists. 64 (2): 141–145.
  • Cormican M, Lowe D, Keane, P, Flynn J, & O’Toole D. (1994). The microbial flora of in-use blood pressure cuffs. Irish Journal of Medical Sciences. 4: 112–113.
  • Fonseca-Reyes, S, de Alba-Garcia J, Parra–Carrillo J, & Paczka-Zapata, J. (2003). Effect of standard cuff on blood pressure readings in patients with obese arms. How frequent are arms of a ‘large circumference’? Blood Pressure Monitoring. 8 (3); 101–106.
  • Geddes L, & Whistler S. (1978). The error in indirect blood pressure measurement with the incorrect size of cuff. American Heart Journal. 96 (1); 4–8.
  • Graves, J. (2001). Prevalence of blood pressure cuff sizes in a referral practice of 430 consecutive adult hypertensives. Blood Pressure Monitoring. 6 (1); 17–20.
  • Haige, M. (1982). Reliability of blood pressure measurement on fat upper arms. Therapetutische Umschau. 39; 87–91.
  • King G. (1967). Errors in clinical measurement of blood pressure in obesity. Clin Sci. 32; 223–237.
  • Kvols, L, Rohlfing, B, & Alexander J. (1969). A comparison of intra–arterial and cuff blood pressure measurements in very obese subjects. Cardiovascular Research Center Bulletin. 7 (3); 118–123.
  • Maxwell M, Waks A, Schroth P, Karam M, & Dornfeld. (1982). Error in blood pressure measurement due to incorrect cuff sizes in obese patients. Lancet. 33–35.
  • Medical care for obese patient: Advice for health care professionals. National task force on the prevention and treatment of obesity. American Family Physician. 65 (1): 81–88.
  • Nielsen P, Larsen B, Holstein P, & Poulsen J. (1983). Accuracy of auscultatory blood pressure measurements in hypertensive and obese subjects. Hypertension. 5 (1); 122–127.
  • Nielsen P, & Janniche H. (1974). The accuracy of auscultatory measurement of arm blood pressure in very obese subjects. Acta Med Scand. 195; 403–409.
  • Pickering, T. (2002). Principles and techniques of blood pressure measurement. Cardiology Clinics. 20; 207–223.
  • Prineas, R. (1991). Measurement of blood pressure in the obese. AEP. 1 (4): 321–336.
  • Stolt M, Sjonell G, Astrom H, Rossner S, & Hansson L. (1993). Improved accuracy of indirect blood pressure measurement in patients with obese arms. AJH. 6; 66–71.
  • Trew C. (1982). Measurement of blood pressure in the obese patient. Anaesthesia & Intensive Care. 10 (4); 374–375.
  • Weeratunga G. (2003). Measuring non-invasive blood pressure in obese patients. Anaesthesia. 58; 597–616